ZOLL CIRCULATION INC AUTOPULSE LI-ION BATTERY; COMPRESSOR, CARDIAC, EXTERNAL Back to Search Results
Model Number 8700-0752
Device Problem Battery Problem (2885)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2019
Event Type  malfunction  
Manufacturer Narrative
Zoll has not received the autopulse li-ion battery for investigation.A supplemental report will be filed when the product is returned and the investigation has been completed.
 
Event Description
During the patient call, the autopulse platform performed the compressions with the fully charged battery, however, the battery capacity was depleted in 15 minutes.The platform powered off without displaying any error message.Per a reporter, the battery was charged 12 hours prior to the placement into the autopulse platform.Another battery was placed into the platform and performed the compressions without any issue.No known impact or consequence to patient information was provided.After the call, the autopulse platform and the same battery were tested with the manikin, the reported issue could not be reproduced by the user.The autopulse platform performed compressions for 45-50 minutes without any problem.See mfr 3010617000-2019-00820 for autopulse platform.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE LI-ION BATTERY
Type of Device
COMPRESSOR, CARDIAC, EXTERNAL
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave
san jose, CA 95131
4085411030
MDR Report Key9093622
MDR Text Key159330907
Report Number3010617000-2019-00821
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001823
UDI-Public00849111001823
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8700-0752
Device Catalogue Number8700-0752-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-