Model Number 1458Q/86 |
Device Problems
Signal Artifact/Noise (1036); Under-Sensing (1661); Device Dislodged or Dislocated (2923)
|
Patient Problem
Twiddlers Syndrome (2114)
|
Event Date 09/06/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results, method, and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During a clinic follow-up, noise was observed on the left ventricular (lv) lead.Diagnostic imaging was performed and revealed lv lead dislodgement due to twiddler's syndrome.The lv lead was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
|
|
Event Description
|
Additional information received that undersensing was noted on the left ventricular (lv) lead.
|
|
Manufacturer Narrative
|
As received, a complete lead was returned in one piece for analysis.The reported events of noise and under-sensing were not confirmed.Electrical testing did not reveal any indication of conductor fractures or internal shorts.Visual inspection and x-ray examination of the lead did not reveal any anomalies.The s-curve hump height was measured within product specification.
|
|