MAUDE Adverse Event Report: ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Signal Artifact/Noise (1036); Under-Sensing (1661); Device Dislodged or Dislocated (2923)
Patient Problem Twiddlers Syndrome (2114)
Event Date 09/06/2019
Event Type  Injury  
Manufacturer Narrative
The results, method, and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During a clinic follow-up, noise was observed on the left ventricular (lv) lead.Diagnostic imaging was performed and revealed lv lead dislodgement due to twiddler's syndrome.The lv lead was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
 
Event Description
Additional information received that undersensing was noted on the left ventricular (lv) lead.
 
Manufacturer Narrative
As received, a complete lead was returned in one piece for analysis.The reported events of noise and under-sensing were not confirmed.Electrical testing did not reveal any indication of conductor fractures or internal shorts.Visual inspection and x-ray examination of the lead did not reveal any anomalies.The s-curve hump height was measured within product specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9099218
MDR Text Key159393493
Report Number2017865-2019-13913
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model Number1458Q/86
Device Lot NumberS000068542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/23/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-