Model Number G141 |
Device Problem
High impedance (1291)
|
Patient Problem
Complete Heart Block (2627)
|
Event Date 07/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
It was reported that a check right ventricular (rv) lead message had been generated for this system.A review of the data identified spikes in the rv pacing impedance measurements with two measurements out of range and greater than 2000 ohms.There were no stored episodes of noise.A boston scientific technical services consultant discussed the clinical observations with the caller who stated they will continue to monitor this patient for potential future issues that could arise with the out of range impedance measurements, oversensing due to interaction with minute ventilation (mv).No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
This supplemental report is being filed to update the investigation conclusion code and additional manufacturing narrative.
|
|