Model Number 176625
Device Problems Mechanics Altered (2984); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, on the third firing of a laparoscopic cholecystectomy, a crack sound was heard and firing was unavailable.Another product was used to complete the case.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with five remaining clips.During functional evaluation, it was observed that the jaws would not close upon actuation of the handle.The instrument body was removed from the shaft and disassembled for visualization of internal components.It was observed that the firing handle linkage was forced through the channel tube lugs, suggesting that the instrument had been fired through the safety interlock.Due to the noted damage, further functional testing of the instrument could not be performed.In addition; the interlock mechanism was triggered as a result of rapidly actuating the handles of the device and or forcing the device, this renders the device inoperable and poses no harm to the patient.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of this condition may occur when an attempt is made to fire the instrument without first loading a clip by using the trigger.The safety interlock feature is designed to prevent the jaws from closing on a vessel in the absence of a loaded clip.If an attempt is made to forcibly fire the instrument while engaged in interlock, the lugs are designed to break as noted and the interlock feature continues to function as intended.No further actions have been deemed necessary at this time.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 18750
DO  18750
MDR Report Key9102214
MDR Text Key159657166
Report Number9612501-2019-01789
Device Sequence Number1
Product Code FZP
UDI-Device Identifier20884521057842
UDI-Public20884521057842
Combination Product (y/n)N
PMA/PMN Number
K954435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2023
Device Model Number176625
Device Catalogue Number176625
Device Lot NumberJ8H2419NX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2019
Date Manufacturer Received10/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-