• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HANDPIECE, ELECTRIC DERMATOME Back to Search Results
Catalog Number 00882100100
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).The device history record for zimmer electric dermatome serial number (b)(4) was reviewed and noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with the device.The record review found no issues with the device and all verifications, inspections, and tests were successfully completed.Using the machine-repair reports and the repair sites folders on (b)(4) as well as sap to query for all repairs on serial number (b)(4) prior to 13 june 2019, the device was noted to have not been previously repaired.Review of the complaint history based on the above keywords found no other similar events related to the reported device; as such, no further additional action is required at this time.The reported event was confirmed by the service technician who performed the evaluation and repair.On 13 june 2019, it was reported during preventative maintenance a dermatome had a defective needle bearing and e-rings and had a motor issue.A zimmer electric dermatome serial number (b)(4) was already at the site and was available for evaluation.Evaluation of the device found that the needle bearing and the e-rings on the dermatome were defective and the motor was found to have been seizing and warm.It was recommended to replace the needle bearing, e-rings, motor, washers, and screws on the device.At the time of the investigation, the device was put on a quotation hold due to the motor seizing during final testing.The device was tested and inspected.While the service technician found that the needle bearing, e-rings, and the motor were all defective, it cannot be determined from the information provided as to what caused the defectives with each of the components.Therefore, a specific root cause of the reported event cannot be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.Review of the information provided during the investigation determined that there are no further actions needed at this time.This complaint will be tracked and trended per complaint trending procedure for any adverse trends that may warrant further action.
 
Event Description
It was reported that during pm, it was found that the defective needle and defective e-rings was replaced and had motor issue.Investigation findings revealed the motor was seizing and warm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key9119018
MDR Text Key160065425
Report Number0001526350-2019-00811
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00882100100
Device Lot Number32325100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2019
Was the Report Sent to FDA? No
Date Manufacturer Received09/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-