TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT 5520; SURGICAL LIGHT Back to Search Results
Model Number 4075520
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 09/09/2019
Event Type  malfunction  
Manufacturer Narrative
A trumpf medical service technician evaluated the light system and found paint chipping on the light head.The light head was replaced and the device is functioning as designed.The investigation identified steps in the preparation for coating process which can lead to lack of paint adhesion resulting in paint chipping.Additionally, collisions of the light heads and use of unapproved cleaners can further contribute to paint chipping from the light system.The damage to the painted surfaces usually does not develop suddenly but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The surface coating process has been updated to include additional testing and verification.Based on this information, no further action is required.
 
Event Description
During an inspection of the light system, chipping paint was found on the light head.No injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT 5520
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key9119248
MDR Text Key195269135
Report Number9681407-2019-00032
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Biomedical Engineer
Remedial Action Replace
Type of Report Initial
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4075520
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-