LIVANOVA DEUTSCHLAND S3 BUBB DET SENSOR, LOW LEVEL II; MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 23-27-40
Device Problem Failure to Sense (1559)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2019
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the sorin s3 bubble detector sensor, low level ii.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported issue.He addressed the failure back to the cable which was found cut.He replaced the level sensor and the problem was solved.Functional verification testing was completed without further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
Livanova (b)(4) received a report that a s3 bubble detector sensor, low level ii was not working during procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 BUBB DET SENSOR, LOW LEVEL II
Type of Device
MONITOR AND/OR CONTROL, LEVEL SENSING, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9122064
MDR Text Key194177575
Report Number9611109-2019-00750
Device Sequence Number1
Product Code DTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K955152
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 09/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number23-27-40
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-