A dispatched service technician could trace the problem back to a malfunction of the power supply and replaced it.The device passed all consecutive tests and is back in use.An in-depth analysis of the replaced power supply was performed in reflection of the fact that it was in operation for 85986 hours.The log files were not available for analysis.The returned power supply unit was installed in a laboratory device.The battery of the primus had a remaining capacity of 95% at the time of power supply replacement.The smell of fire described by the user could be reproduced and is attributable to the heat conductive paste used in this ten years old power supply unit.An oily substance formed, which could lead to failures of some components.In the event of a power supply failure, the device continues to operate using the internal buffer batteries without limiting functionality.At the same time a power-supply error alarm is issued.The remaining capacity of the batteries is displayed to the user.In addition, a warning message is sent to the user if the battery capacity falls below the 20% and 10% limits.Manual ventilation with the built-in breathing bag remains possible.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|