Catalog Number 03260763160 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reporter's product was requested for investigation and replacement product was sent to the reporter.
|
|
Event Description
|
The initial reporter stated they received discrepant nitrite results for three patient samples tested with chemstrip 10 md urine test strips on urisys 1100 analyzer serial number (b)(4).One patient sample tested on (b)(6) 2019 resulted with a positive nitrite value when the test strip is tested on the urisys 1100 analyzer.When reading the same test strip visually, the nitrite value appears to be negative.Two patient samples tested on (b)(6) 2019 resulted with a positive nitrite value when the test strip is tested on the urisys 1100 analyzer.When reading the same test strips visually, the nitrite values appeared to be negative.The reporter tried testing water with the same test strips twice and each time, the water had a positive nitrite value when the test strip was measured with the urisys 1100 analyzer.The test strips appeared to have a negative nitrite values when read visually.The reporter opened a new vial of the same test strip lot number and tried testing the water again.Both the analyzer measurement and visual reading of the test strip had negative nitrite values.
|
|
Manufacturer Narrative
|
The customer sent the u1100 analyzer s/n ux09662560 and one vial of chemstrip 10 md urine test strips lot 35703302 with 8 of 100 test strips remaining for further investigation.The returned customer strips were discolored and the lid of the vial showed signs of humidity.The issue is consistent with the vial not being closed correctly or being opened too long.The retention material shows no abnormalities.
|
|
Manufacturer Narrative
|
The customer's urisys 1100 analyzer was clean and showed no damages.The customer's urisys 1100 analyzer was measured with retention test strip lot number 43065200 with native urine.The analyzer showed no false positive results and fulfilled requirements.
|
|
Search Alerts/Recalls
|