• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH SYNFLATE BALLOON/LARGE-STERILE; POLYMETHYLMETHACRYLATE BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH SYNFLATE BALLOON/LARGE-STERILE; POLYMETHYLMETHACRYLATE BONE CEMENT Back to Search Results
Catalog Number 03.804.702S
Device Problem Deflation Problem (1149)
Patient Problem No Code Available (3191)
Event Date 08/29/2019
Event Type  Injury  
Manufacturer Narrative
Additional device product code: hrx.Complainant part is not expected to be returned for manufacturer review / investigation.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and / or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that during an unknown procedure on (b)(6) 2019 with the synflate system, it was impossible to deflate three (3) synflate l balloons in the vertebra.It was not possible to remove the synflate balloons with the working trocar.The procedure was completed by pulling the not completely deflated balloon to the top of the working trocar (3x) and completely removing both the trocar and the balloon.Procedure was delayed by thirty (30) minutes.Procedure was completed successfully by using same like product.Patient outcome is reported as stable.Concomitant device reported: balloon inflation system (part # unknown, lot # unknown, quantity 1); unknown torcar (part # unknown, lot # unknown, quantity 3).This report is for one (1) synflate balloon / large-sterile.This is report 2 of 3 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part number: 03.804.702s, lot number: 0818005, manufacturing site: selzach, supplier: (b)(4), release to warehouse date: 17.Sep.2018, expiry date: 01.Aug.2020.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNFLATE BALLOON/LARGE-STERILE
Type of Device
POLYMETHYLMETHACRYLATE BONE CEMENT
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key9136841
MDR Text Key163162605
Report Number8030965-2019-68860
Device Sequence Number1
Product Code NDN
UDI-Device Identifier07611819525405
UDI-Public(01)07611819525405
Combination Product (y/n)N
PMA/PMN Number
K130146
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2020
Device Catalogue Number03.804.702S
Device Lot Number0818005
Was Device Available for Evaluation? No
Date Manufacturer Received10/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNK - INSERTION INSTRUMENTS: TROCAR: SPINE; UNK - SYNFLATE/VBS
Patient Outcome(s) Required Intervention;
-
-