The reported event was unconfirmed, since the reported failure could not be reproduced.The product was used for treatment purposes.The device met specifications.Visual evaluation of the silicone noted one opened silicone foley with balloon inflated without original packaging.Visual inspection of the sample noted no obvious defects.The 7cc solution was removed from catheter.The catheter balloon was re-inflated with 10 ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) balloon concentricity was observed to be 70:30.The balloon rested for 30 minutes without leaks and passively deflated without issue or cuffing, returning 10ml of solution.The drainage lumen was flushed with methylene blue solution (3 drops 1% aq methylene blue per 100ml of distilled water) with no leaks.Active length of the catheter balloon was measured ( 0.6990") and found to be within specification (0.6-0.9 inches).The catheter was confirmed to be size 14 fr.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿recommended inflation capacities 3cc balloon: use 5cc sterile water.5cc balloon: use 10cc sterile water.30cc balloon: use 35cc sterile water.Do not exceed recommended capacities.Note: aggressive traction, particularly in the presence of suturing, is not recommended for silicone foley catheters." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|