As reported in literature finding, the following adverse event occurred during a case study for lead extraction procedures which took place between october 2011 and december 2018."in dcl group there were 2 major complications (cardiac tamponades requiring surgical intervention).During extraction of dc icd leads (dual-coil implantable cardioverter) with telescopic sheaths (cook lead extraction byrd polypropylene extra long dilator sheath set, lr-pplbes-13.0-xl) the free wall of the right ventricle was damaged and caused tamponade.Both complications were caused by a lead tip penetration into the myocardium without lead perforation in tte (transthoracic echocardiography) and cardiac ct (computed tomography) angiogram." the adverse event sited in this excerpt, which this report is intended to address, is indicated to be a cardiac tamponade, which required surgical intervention.No other consequences to the patient have been reported as occurring.At this time, no additional information is available.
|
Additional information: h6- ec method code desc - 1: device not returned (4114).H6- ec results code desc - 1: no findings available (3221).H6- ec conclusions code desc - 1: known inherent risk of device (22).Investigation-evaluation: as per journal article: effectiveness and safety of transvenous extraction of single- versus dual-coil implantable cardioverter-defibrillator leads at single-center experience: the aim of the study was to assess the effectiveness and safety of single- and dual-coil icd leads transvenous extraction using mechanical systems.We performed transvenous lead extraction (tle) of 197 icd leads in 196 patients.There were 46 (23.3%) dual-coil leads removed from 46 (23.5%) patients.Adhesions developing around the svc coil add to the difficulty of extraction of icd lead if lead removal is required.The selected groups of patients (scl and dcl) were compared by assessing demographic data (age, sex), body mass index (bmi), the new york heart association (nyha) functional classification and left ventricular ejection fraction (lvef), indications for icd implantation (primary or secondary prevention), comorbidities including diabetes mellitus and coronary artery disease, laboratory studies (hemoglobin concentration, creatinine level, estimated glomerular filtration rate [egfr]), types of implanted cied, number of cied-related procedures before tle (implantation, reimplantation, device upgrade), indications for tle.For each electrode removed, the efficiency according to the ehra consensus was determined: complete lead removal--lead explant or extraction with removal of all targeted lead material, incomplete lead removal--lead explant or extraction where part of the lead remains in the patient¿s body (vascular or extravascular).All the procedures were performed in hybrid operating room with on-site cardiothoracic surgical standby, under general anesthesia or intravenous sedation.While extracting the lead we used a stepwise approach.Simple traction was the first extraction technique used in active fixation and relatively young leads.At the next step we utilized nonpowered extraction tools.Standard stylets were used to stiffen the leads.If the lumen of the lead allowed the insertion of the stylet to its tip, the locking stylets (liberator beacon tip locking stylet, cook medical or lead locking device, spectranetics) were used from december 2013.We utilized telescoping dilators (cook medical, bloomington, in).Powered extraction tools (cutting sheath system--evolution, cook medical) were usually applied when previously used nonpowered extraction tools failed.In case of leads which were completely in the cardiovascular system either before the procedure or as a result of lead fracture during the procedure, the femoral approach was utilized.Laser techniques and electrosurgical sheaths were not used.As stated in literature finding, "in dcl group there were 2 major complications (cardiac tamponades requiring surgical intervention).During extraction of dc icd leads with telescopic sheaths the free wall of the right ventricle was damaged and caused tamponade.Both complications were caused by a lead tip penetration into the myocardium without lead perforation in tte and cardiac ct angiogram.Additional information on major complication in both groups was presented in table 3." (of the journal article attached to this complaint).Per ifu (d00078677 rev002): "weigh the relative risks and benefits of intravascular catheter/lead removal procedures in cases when: the item to be removed is of a dangerous shape or configuration, the likelihood of catheter/lead disintegration resulting in fragment embolism is high or vegetations are attached directly to the catheter/lead body.Catheter/lead removal devices should be used only at institutions with thoracic surgical capabilities." "catheter/lead removal devices should be used only by physicians knowledgeable in the techniques and devices for catheter/lead removal." "potential adverse events related to the procedure of intravascular extraction of catheters/leads include (listed in order of increasing potential effect): dislodging or damaging nontargeted catheter/lead, chest wall hematoma, thrombosis, arrhythmias, acute bacteremia, acute hypotension, pneumothorax, stroke, migrating fragment from catheter/object, pulmonary embolism, laceration or tearing of vascular structures or the myocardium, hemopericardium/pericardial effusion, cardiac tamponade, hemothorax, cardiac arrest, death." this report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|