Results: the cat6 was ovalized from approximately 124.0 ¿ 129.0 cm from the hub.The total length of the cat6 was approximately 135.5 cm.Conclusions: evaluation of the returned cat6 revealed ovalizations on the distal shaft.If the cat6 is forcefully pinched or gripped prior to insertion into a parent device, damage such as ovalizations may occur.This damage may have contributed to the sudden stoppage of aspiration during the procedure.During functional testing, the cat6 was connected to a demonstration pump max and was able to aspirate liquid without an issue.The cat6 total length was measured and was within specification; therefore, the reported stretch could not be confirmed.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|