Catalog Number 71158020
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2019
Event Type  malfunction  
Event Description
It was reported that during surgery the drad packs were opened in theatre.The devices were found with rust on the screwdriver handles.The surgeon used a competitor device to finish the surgery.No further information available yet.
 
Manufacturer Narrative
The devices, used in treatment, was returned for evaluation.A visual inspection was performed and confirmed the stated failure.The metal tips of the drivers showed signs of rust.The devices were manufactured in 2014 and 2015.Upon review of the returned parts it was noted that the d-rad t7 capture driver hex bit is corroded while still packaged.Nothing noted in the dhr or internal operations that would indicate this issue to be caused internally.Our lab noted: the presence of rust in several sections of the devices.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A root cause could not be determined with the information provided.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
D-RAD SMART PACK 4H LEFT WIDE PLATE
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9147996
MDR Text Key161002891
Report Number1020279-2019-03541
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
PMA/PMN Number
K132296
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71158020
Device Lot Number15EM18790
Date Manufacturer Received10/19/2020
Is This a Reprocessed and Reused Single-Use Device? No