Catalog Number 07026935190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).(b)(4).
 
Event Description
The initial reporter complained of questionable thyroid results for 1 patient sample on a cobas 8000 e 801 module compared to competitor methods.The initial results from the customer's e 801 were reported outside of the laboratory.The competitor methods were considered correct.There were discrepant results for elecsys ft3 iii, elecsys prolactin assay, elecsys estradiol assay, elecsys testosterone ii assay from the customer's e 801 module compared the siemens centaur method.There were discrepant results for elecsys tsh assay from the customer's e 801 module compared to the siemens centaur and beckman methods.There were discrepant results for roche diagnostics cobas elecsys anti-tpo from the customer's e 801 module compared to the beckman method.There were discrepant results for elecsys dhea-s from the customer's e 801 module compared to the siemens immulite 2000 method.This medwatch will cover anti-tpo.Refer to medwatch with patient identifier (b)(6) for information on the tsh results, medwatch with patient identifier (b)(6) for information on the prolactin results, medwatch with patient identifier (b)(6) for information on the estradiol results, medwatch with patient identifier (b)(6) for information on the testosterone results, medwatch with patient identifier (b)(6) for information on the ft3 iii results, and medwatch with patient identifier (b)(6) for information on the dhea-s results.Refer to the attachment to the medwatch for all patient data.The customer's cobas e 602 module serial number was not provided.
 
Manufacturer Narrative
One patient sample was returned for investigation.The investigation identified an interfering factor against the streptavidin component of the reagent in the sample.The interfering factor is documented in product labeling which states: interference in the method sheets of all applicable products: in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO
Type of Device
SYSTEM, TEST, THYROID AUTOANTIBODY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key9151866
MDR Text Key218206261
Report Number1823260-2019-03549
Device Sequence Number1
Product Code JZO
Combination Product (y/n)N
PMA/PMN Number
K051890
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Catalogue Number07026935190
Device Lot Number39837801
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
L-THYROXIN
Patient Age26 YR
-
-