Manufacturer Narrative
Date of event is an estimated date based off the aware date as the exact event date was not reported.Device evaluated by manufacturer: the device was returned for analysis.The catheter returned has the hub detached (the heat shrink was received properly attached to the hub); moreover, the flare of the extrusion was found formed, it is evidence that the manufacturing process was properly performed.The catheter returned with detached hub.The catheter was also cut at 12cm from the flare end.The rest of the catheter was not returned for analysis.
 
Event Description
Reportable based on device analysis completed on 12sep2019.A flexima apdl catheter drainage was received with no reported issues.However, analysis revealed hub detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA APDL
Type of Device
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9152291
MDR Text Key161712696
Report Number2134265-2019-11948
Device Sequence Number1
Product Code FFA
UDI-Device Identifier08714729323143
UDI-Public08714729323143
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K944290
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/15/2022
Device Model Number30935
Device Catalogue Number30935
Device Lot Number0023803685
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-