• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT Back to Search Results
Model Number G50697
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hemostasis (1895)
Event Date 09/09/2019
Event Type  Injury  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The size of tissue sample or biopsy to be excised can be controlled by how the user handles the forceps.If excessive pressure was applied during advancement into the tissue or during handle manipulation, this could have contributed to the reported observation.The instructions for use (ifu) direct the user to, ¿advance forceps to biopsy or retrieval site, then open cups and advance into tissue to be biopsied or objected to be retrieved.Using slight pressure on handle, close forceps around tissue or object.Maintain gentle handle pressure to keep cups closed and gently withdraw forceps from site.¿ the ifu warns, "these single-use biopsy forceps should only be used to biopsy tissue where possible bleeding or hemorrhage will not present a danger for patients.Adequate plans for management of potential bleeding or hemorrhage and appropriate airway management should be in place." the ifu lists potential complications as, "those associated with gastrointestinal endoscopy include but are not limited to: perforation, bleeding or hemorrhage, aspiration, fever, infection, allergic reaction to medication, hypotension, respiratory depression or arrest, cardiac arrhythmia or arrest." prior to distribution, all captura pro biopsy forceps are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an unusual occurrence.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During a colonoscopy, the physician used a cook captura pro biopsy forceps with spike.After the biopsy there was a large hematoma and some slight bleeding at the site.The physician used a cook instinct endoscopic hemoclip to stop the bleeding.No further action was needed due to this occurrence.The physician confirmed there was no harm to the patient.A section of the device did not remain inside the patient¿s body.The patient required placement of a hemoclip to control the bleeding due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE
Type of Device
PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key9156485
MDR Text Key161774873
Report Number1037905-2019-00580
Device Sequence Number1
Product Code PTS
UDI-Device Identifier00827002506978
UDI-Public(01)00827002506978(17)220624(10)W4229969
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/24/2022
Device Model NumberG50697
Device Catalogue NumberBF-2.4SL-230SP-40
Device Lot NumberW4229969
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/09/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/24/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-