• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC KINCISE CUP-ADAPTER-PINNACLE STRAIGHT; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC KINCISE CUP-ADAPTER-PINNACLE STRAIGHT; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 1011-01-101
Device Problems Crack (1135); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reporter address was not provided.The device manufacture date was unknown.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported that during an unspecified surgical procedure, while implantation of the acetabular liner, it was observed that the adapter device broke.According to the reporter, the device did not break in pieces, it was broken at the t handle junction and bent but is still in one piece.The device is no longer functional.It was not reported if there were delays in the surgical procedure or if a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.Quality engineering evaluated the device and it was determined that the t-handle was broken.Therefore, the reported condition was confirmed.The assignable root cause was traced to the user most likely from side loading when used with the impactor which is not what the device was designed for, or being dropped.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The date of manufacture was reported as unknown in the initial report and has been updated to 06/18/2018.The actual device has been returned and is currently pending evaluation.Once the investigation has been completed, a supplemental medwatch report will be submitted accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINCISE CUP-ADAPTER-PINNACLE STRAIGHT
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
MDR Report Key9162490
MDR Text Key161809068
Report Number1045834-2019-54648
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00850915006099
UDI-Public850915006099
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1011-01-101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2019
Date Manufacturer Received10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-