Model Number PM2240
Device Problems Pacemaker Found in Back-Up Mode (1440); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/25/2019
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It a was reported that the asymptomatic patient presented in clinic during follow-up in backup vvi.A firmware download was performed, and the device was restored to normal function.The device remains implanted.The patient is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY RF DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9163393
MDR Text Key161442979
Report Number2938836-2019-14420
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734507073
UDI-Public05414734507073
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2017
Device Model NumberPM2240
Device Lot NumberA000023308
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/08/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Weight83
-
-