SMITH & NEPHEW MEDICAL LTD. ALLEVYN LIFE SACRUM (SMALL) PACK OF 10; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Catalog Number 66801306
Device Problem Insufficient Information (3190)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 01/01/1901
Event Type  malfunction  
Event Description
It was reported that the product is "shedding" silicone onto the patient upon removal.It is difficult to remove from the skin and residues were left inside the wound bed.Customer is finding some dressings where the silicone is pulling off when the plastic backing is removed.Customer reported that this happened many times, however, an exact number is not available.
 
Manufacturer Narrative
We have now concluded our investigation into this complaint.The returned sample was evaluated which confirmed the presence of silicone offsetting, however, the wound contact surface of allevyn life sacrum is coated with a gentle silicone adhesive layer that ensures non-traumatic removal at dressing changes.The silicone adhesive will feel sticky when compared to an acrylic adhesive and is specially designed to be very soft and gentle, but can soften further, particularly at higher temperatures.This is an inherent characteristic of all of the silicone adhesives and gives them their soft / gentle properties.It is therefore possible if the product has been stored at a high temperature, this could have contributed to the reported issue.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.We will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN LIFE SACRUM (SMALL) PACK OF 10
Type of Device
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull
UK 
MDR Report Key9164349
MDR Text Key161496212
Report Number8043484-2019-00699
Device Sequence Number1
Product Code NAD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 11/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801306
Device Lot Number201916
Date Manufacturer Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No