Visual inspection: the protege rx was removed from the packaging and inspected.The tuohy-borst valve was tight.No external damages to the protege rx were noted.Approximately 0.3cm of the stent was exposed outside of the distal portion of the catheter shaft.Under microscope dried blood was visible at the proximal end of the loaded stent at the retainer.The approximate length the loaded stent was 3cm.Dried blood was observed within the catheter shaft and between struts of the stent.Functional testing: a 0.014¿ guidewire from the lab was front loaded the catheter after the deployment system was flushed with water.The protege rx was loaded the deployment apparatus from within the lab.3.11 lbs of force was applied to the deployment grip, but the stent showed no signs of deployment.If information is provided in the future, a supplemental report will be issued.
|