• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION TMJ SYSTEM CROSS DRIVE FOSSA SCREW; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION TMJ SYSTEM CROSS DRIVE FOSSA SCREW; PLATE, BONE Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cancer (3262)
Event Date 06/19/2019
Event Type  Injury  
Manufacturer Narrative
Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.It is unknown at this time if the device will be returned for evaluation.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00440, 0001032347-2019-00441, 0001032347-2019-00442.Concomitant medical products¿ tmj system left fossa component, small, cat# 24-6563, lot# 166460a; tmj system left standard mandibular component, cat# 24-6556, lot# 831890b; "2,4mm" system high torque (ht) cross-drive screw, cat# 91-2710, lot# unk; tmj system cross drive fossa screw, cat# 99-6577, lot# unk.Occupation ¿ patient.
 
Event Description
It was reported a patient underwent a revision to remove temporomandibular implants on the left side.The implant was removed due to squamous cell carcinoma in the oral cavity.Attempts have been made but no further information is available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is non-verifiable.No product was returned; therefore, no functional tests or inspections could be performed.An x-ray was received, however, the x-ray showed a right mandible plate which spanned a gap between the patient's ramus and chin, and was fractured at the chin.Five follow up attempts were made for additional information regarding this implant's identity and scans relating to the left mandible and left fossa components in this complaint file.Tmj tracking confirmed that a left mandible (part# 24-6556, lot# 831890b), left fossa (part# 24-6563, lot# 166460a) and screws (part# 91-2710, 99-6577) were implanted on (b)(6) 2010.The provided x-ray was taken on (b)(6) 2019, prior to removal of the right mandible plate, and a nurse from the hospital confirmed that the right mandible was removed (b)(6) 2019.The x-ray was reviewed by the clinical research team, who concluded there was no evidence indicating that a tmj replacement surgery ever occurred on either side of the patient in the images.The most likely underlying cause cannot be determined.There are no additional indications of manufacturing defects.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ SYSTEM CROSS DRIVE FOSSA SCREW
Type of Device
PLATE, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9173185
MDR Text Key161743515
Report Number0001032347-2019-00443
Device Sequence Number1
Product Code JEY
UDI-Device Identifier00841036057442
UDI-Public00841036057442
Combination Product (y/n)N
PMA/PMN Number
K910038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number99-6577
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-