• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK FEMALE EXTERNAL Back to Search Results
Catalog Number PWF030F
Device Problem Leak/Splash (1354)
Patient Problem Urinary Tract Infection (2120)
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation.A potential root cause could be incorrect device placement.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "contraindications: patients with urinary retention.Warnings: do not use purewick female external catheter with bedpan or any material that does not allow for sufficient airflow.To avoid potential skin injury, never push or pull the purewick female external catheter against the skin during placement or removal.Never insert the purewick female external catheter into vagina, and anal canal or other body cavities.Discontinue use if an allergic reaction occurs.After use, this product may be a potential biohazard.Dispose of in accordance with applicable local, state and federal laws and regulations.Precautions: not recommended for patients who are: -agitated, combative or uncooperative and might remove the purewick female external catheter.Having frequent episodes of bowel incontinence without a fecal management system in place.Experiencing skin irritation or breakdown at the site.Experiencing moderate/heavy menstruation and cannot use a tampon.Do not use barrier cream on the perineum when using the purewick female external catheter.Barrier cream may impede suction.Not recommended for use on patients who have undergone recent surgery of the external urogenital tract.Always assess skin for compromise nd perform perineal care prior to placement of a new purewick female external catheter.Maintain suction until the purewick female external catheter is fully removed from he patient to avoid urine backflow.Recommendations: perform each step with clean technique.In the home setting, wash hands thoroughly before device placement.Prior to connecting the purewick female external catheter to hospital wall suction tubing, verify suction function by covering the open end of the suction tubing with one hand and observing the pressure dial.If the pressure does not increase when the line is covered, verify that the tubing is secured, connected and not kinked.Ensure the purewick female external catheter remains in the correct position after turning the patient.Remove the purewick female external catheter prior to ambulation.Properly placing the purewick female external catheter snugly between the labia and gluteus holds the purewick female external catheter in place for most patients.Mesh underwear may be useful for securing the purewick female external catheter for some patients.Assess device placement and patient's skin at least every 2 hours.Replace the purewick female external catheter every 8-12 hours or when soiled with feces or blood.Change suction tubing per hospital protocol or at least every thirty (30) days.".
 
Event Description
It was reported that the patient's daughter said her mother was waking up wet and constantly getting utis, one after another.It was unknown if the patient received medication for the uti.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK FEMALE EXTERNAL
Type of Device
PUREWICK
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9176423
MDR Text Key161965427
Report Number1018233-2019-06437
Device Sequence Number1
Product Code NZU
UDI-Device Identifier00801741163081
UDI-Public(01)00801741163081
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberPWF030F
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/20/2019
Initial Date FDA Received10/10/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-