LIMACORPORATE SPA SMR -REVERSE RESECTION JIG DPA; SMR -REVERSE RESECTION JIG DPA (KWT, HSD ) Back to Search Results
Model Number 9013.52.304
Device Problem Entrapment of Device (1212)
Patient Problem No Code Available (3191)
Event Date 09/20/2019
Event Type  malfunction  
Manufacturer Narrative
By checking the dhr of the lot #, no anomaly was found.We will submit a final mdr once the investigation will be concluded.
 
Event Description
Intra-operative issue occurred on (b)(6) 2019.During smr primary surgery the pin got stuck into the smr -reverse resection jig code (b)(4), lot #14aa458.According to the information reported, the pin was manufactured by limacorporate.Consequently, the surgeon performed the humeral cut by freehand.The event prolonged the surgery of 3 minutes.Event occurred in (b)(6).
 
Event Description
Intra-operative issue occurred on (b)(6) 2019.During smr primary surgery the pin got stuck into the smr -reverse resection jig code 9013.52.304 lot #14aa458.Consequently, the surgeon performed the humeral cut by freehand.According to the information reported, the pin (code and lot# unknown) was manufactured by limacorporate.The event prolonged the surgery of 3 minutes.Event happened in australia.
 
Manufacturer Narrative
Check of the dhr: by checking the dhr of the lot #14aa458, no pre-existing anomaly was detected on 56 smr-reverse resection jigs manufactured.Therefore, all the items were up to specifications.Instrument analysis: the instruments involved in the intra-operative incident (resection jig and pin) were not returned to limacorporate for technical investigation.Picture of the instrument shows the pin jammed into the resection jig.However, based on the provided picture, it is not possible to do any spaculation on the root cause of the event.Based on the few information received, we cannot investigate the root cause of the incident.However, considering that the instrument was 5 years old when the incident occurred and considering the presence of similar complaints on the same instrument code, we can hypothesize that the event was due to a combination of wear due to usage, design factor and surgical factor.Corrective action has already been implemented on family of code 9013.52.304/305.After receiving previous similar complaints, in may 2016 the technical drawing of the smr reverse resection jig was adjusted to slightly increase the holes diameter of the jig and therefore further reduce the intra-operative risk of seizure of the pin into the jig holes.Pms data: according to our pms data and considering the resection jigs code 9013.50.304 and 9013.52.304-305 manufactured prior drawing improvement, (b)(4).Since the new design has been introduced, (b)(4).Both these occurrence rates are overestimated because they do not consider the reuse of the instrument.Most of these issues happened after several uses of the resection jigs or of the pins.None of these cases caused serious consequences on the patient or prolonged surgical time.Limacorporate will continue to monitor the market to promptly detect the possible recurrence of this issue and assess the effectiveness of the corrective action performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR -REVERSE RESECTION JIG DPA
Type of Device
SMR -REVERSE RESECTION JIG DPA (KWT, HSD )
Manufacturer (Section D)
LIMACORPORATE SPA
via nazionale 52
villanova di san daniele, udine 33038
IT  33038
MDR Report Key9179933
MDR Text Key194472278
Report Number3008021110-2019-00115
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K100858
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number9013.52.304
Device Lot Number14AA458
Date Manufacturer Received10/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-