The initial reporter stated they received discrepant results for two samples from the same patient tested with phos2 phosphate (inorganic) ver.2 on a two cobas 8000 c 502 module analyzers and an unknown roche cobas clinical chemistry analyzer.No incorrect results were reported outside of the laboratory.The first sample initially resulted with a phosphorus value of 7.65 mg/dl when tested on the first c 502 analyzer.The sample was repeated on an unknown competitor dry chemistry system, resulting with a value of 3.1 mg/dl on (b)(6) 2019.The sample was repeated on an unknown roche clinical chemistry analyzer, resulting with a value of 8.16 mg/dl on (b)(6) 2019.The sample was also repeated on a second c 502 analyzer, resulting with a value of 8.0 mg/dl on (b)(6) 2019.The second sample initially resulted with a phosphorus value of 4.63 mg/dl when tested on the first c 502 analyzer on (b)(6) 2019.The sample was repeated on an unknown competitor dry chemistry system, resulting with a value of 2.6 mg/dl.The sample was also repeated on a second c 502 analyzer, resulting with a value of 4.54 mg/dl on (b)(6) 2019.The serial number of the first c 502 analyzer was (b)(4).The serial numbers of the second c 502 analyzer and unknown roche clinical chemistry analyzer were requested, but not provided.
|