MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH MAQUET HCU 40; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH MAQUET HCU 40; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number HEATER COOLER UNIT
Device Problems Electrical /Electronic Property Problem (1198); Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/06/2019
Event Type  malfunction  
Manufacturer Narrative
The error "mains power fail" could be detected by the getinge service technician when analysing the device log data.The repair of the device will be done by trained customer.Since no parts have been replaced by field service technician (fst) of getinge, there is no most probable root cause be determined.The failure is covered in the current risk analysis (v17).Thus the reported failure could be confirmed.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopumonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint number (b)(4).It was reported that the hcu 40 mains power fails during perfusion (treatment).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET HCU 40
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key9190979
MDR Text Key199893983
Report Number8010762-2019-00327
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K130300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHEATER COOLER UNIT
Device Catalogue NumberHCU 40, 701044054
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/07/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-