It was reported that stent dislodgment occurred.The 100% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vessel.A 6.0mm x 18mm x 150cm express sd renal/biliary stent was advanced to treat the lesion.However, upon inserting the device, the stent came off the balloon and became stuck in the patient's femoral artery.The physician attempted to pull the stent out with a different balloon but was unsuccessful.The or team was then called to perform a cutdown to remove the stent.No further patient complications were reported and the patient was stable.
|
Device evaluated by mfr.: returned product consisted of 55cm of the distal shaft, hypotube, stent, balloon and tip of an express sd catheter.The proximal end of the hypotube and hub were not returned.The balloon is loosely folded.The hypotube, outer shaft, inner shaft, balloon, stent and tip were visually and microscopically examined.The hypotube is separated 55cm from the distal tip.There are numerous hypotube and shaft kinks.The stent is pushed proximally on the shaft and is severely damaged.The tip is damaged.Inspection of the remainder of the device and stent presented no other damage or irregularities.
|
It was reported that stent dislodgment occurred.The 100% stenosed target lesion was located in the mildly tortuous and severely calcified iliac vessel.A 6.0mm x 18mm x 150cm express sd renal/biliary stent was advanced to treat the lesion.However, upon inserting the device, the stent came off the balloon and became stuck in the patient's femoral artery.The physician attempted to pull the stent out with a different balloon but was unsuccessful.The or team was then called to perform a cutdown to remove the stent.No further patient complications were reported and the patient was stable.
|