• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 701046412 - ROTAFLOW EN/NORTH AM US-PLUG
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
A follow up medwatch will be submitted when additional informations bekomes available.Device not returned not eval.
 
Event Description
Rotaflow: head error greater than 1500 rpm occurred when connecting with another rotaflow during priming.Complaint# (b)(4).
 
Manufacturer Narrative
2020-01-24: the concerned drive and console was received by maquet under rma#(b)(4).2020-02-18: the failure head error could be confirmed after testing in the service department.Same failure was already investigated at emtec report no.Rma2020-10009 on 2020-02-24 (complaint lot 267240; rma#(b)(4)): the defective drive was sent to the supplier em tec for further root cause investigation: 2020-02-24: em tec report no.Rma2020-10009: the reported head error is a result of wrong handling of the user see below causes.The reported failure could be reproduced and confirmed.Most possible root cause could be determined as: 1.The head error is caused by the hot plug.When device is in operation and the power plug is plugged in or out.And this leads to a damage at the control board of the rotaflow.2.The head error follows by the sig error.This is when the ultrasonic crème is applied to the flow bubble sensor.Then the centrifugal pump is causing backflow and this leads to the head error.3.The head error is also caused by shaking the drive.This is when the motor (which is controlled by the optical tacho) is not blocked when adjusting to 0 then it could lead to error when slight shaking.The motor could then slightly rotate.4.The head error can be caused by connection issues between the console (rfc) and the drive (rfd).This is when the cable connection is disturbed by defective pins.Furthermore, the instructions for use of the rotaflow system, see rotaflow system user manuel, mcv-ga-10000703-de-11 contain detailed descriptions to prevent an ¿error head¿.The occurrence rate was calculated for the reported failure and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
Complaint#(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
MDR Report Key9202448
MDR Text Key194538300
Report Number8010762-2019-00331
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 04/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701046412 - ROTAFLOW EN/NORTH AM US-PLUG
Device Catalogue Number701046412
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2020
Date Manufacturer Received04/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-