Model Number CD3369-40Q
Device Problems Signal Artifact/Noise (1036); High impedance (1291); No Device Output (1435); Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented at a follow-up in clinic post-implant procedure.Upon interrogation, the right ventricular channel on the implantable cardioverter defibrillator exhibited high pacing impedance, episodes noise, episodes of failure to sense, and inconsistent capture.The physician unattached and reattached the right ventricular lead on (b)(6) 2019, which resolved the issue.The physician alleged that an air pocket in the device header caused the electrical anomalies.The patient experienced no adverse consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key9205561
MDR Text Key162814895
Report Number2017865-2019-15235
Device Sequence Number1
Product Code LWS
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model NumberCD3369-40Q
Device Lot NumberA000085123
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/08/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-