MEDTRONIC NEUROMODULATION SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97702 |
Device Problems
Migration or Expulsion of Device (1395); Human-Device Interface Problem (2949)
|
Patient Problem
Therapeutic Effects, Unexpected (2099)
|
Event Date 08/15/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2016, product type: lead.Product id: 977a260, serial#: (b)(4), implanted: (b)(6) 2016, product type: lead.Product id: 977a260, serial/lot #: (b)(4), ubd: 27-jul-2020, udi#: (b)(4).Product id: 977a260, serial/lot #: (b)(4), ubd: 08-jul-2020, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from the manufacturer's representative regarding the patient who was implanted with a neurostimulator for spinal cord stimulation - spinal pain.It was reported that the patient experienced a change in her stimulation coverage abruptly.Patient reported increased activity.X-ray and impedances tested.X-ray confirmed a migration.Lead revision planned but not scheduled.No further complications were reported/anticipated.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the rep.It was reported that the cause of migration was a failure of the original injex anchors.At replacement it was determined by the physician they were not deployed correctly.The physician was aware.
|
|