MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK ADVANTUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS CLINITEK ADVANTUS Back to Search Results
Catalog Number 10636672
Device Problem Non Reproducible Results (4029)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/23/2019
Event Type  malfunction  
Manufacturer Narrative
The customer stated that repeat testing was performed to confirm correct results and a corrected report was issued.The customer declined to send any materials for further investigation.Proper technique and maintenance were reviewed with the customer as well as possible causes of a falsely depressed leukocyte result.The customer had not lubricated the push bar as often as recommended and are operating at a slightly less than recommended temperature.Since they installed the new fixed platform and lubricated the push-bar rail, they have not had any further discrepancies.The instrument is operational.
 
Event Description
The customer reported discrepant leukocyte results on the clinitek advantus when compared to the microscopic examination of the urine sediment.There is no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK ADVANTUS
Type of Device
CLINITEK ADVANTUS
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
KIMBALL ELECTRONICS
ul. pozenanska 1/c
poland sp.z o.o
tarnowo podgorne, 62080
PL   62080
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key9215846
MDR Text Key164915883
Report Number3002637618-2019-00113
Device Sequence Number1
Product Code JIL
UDI-Device Identifier00630414593562
UDI-Public00630414593562
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063276
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 10/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10636672
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/23/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age20 YR
-
-