LIMACORPORATE SPA SMR - ANATOMIC RESECTION JIG; SMR - ANATOMIC RESECTION JIG (KWT, HSD) Back to Search Results
Model Number 9013.50.304
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
By checking the manufacturing chart of the lot involved, no anomaly was found.This is the first and only complaint received on this lot#.We will submit a final mdr once the investigation will be concluded.
 
Event Description
Intra-operative issue occurred on (b)(6) 2019.During anatomic smr surgery, the pin got stuck into the smr anatomic resction jig code (b)(4), lot #15ab119.Cutting jig from another tray was available and used.Surgery was prolonged of less than 10 minutes.The event happened in (b)(6).
 
Manufacturer Narrative
Check of the dhr: by checking the dhr of the lot #15ab119, no pre-existing anomaly was detected on 31 smr anatomic resection jigs manufactured.This is the first and only complaint received on this lot#.Instrument analysis: the instruments involved in the intra-operative incident (resection jig and pin) were not returned to limacorporate for technical investigation.Picture of the instrument shows the pin jammed into the resection jig.However, based on the provided picture, it is not possible to do any spaculation on the root cause of the event.Based on the few information received, we cannot investigate the root cause of the incident.No further complaints were received on this lot#, therefore we cannot judged this case as product related.However, considering that the instrument was 4 years old when the incident occurred and considering the presence of similar complaints on the same instrument code, we can hypothesize that the event was due to a combination of wear due to usage, design factor and surgical factor.Corrective action has already been implemented on family of code 9013.52.304/305.After receiving previous similar complaints, in may 2016 the technical drawing of the smr reverse resection jig was adjusted to slightly increase the holes diameter of the jig and therefore further reduce the intra-operative risk of seizure of the pin into the jig holes.Pms data: according to our pms data and considering the resection jigs code 9013.50.304 and 9013.52.304-305 manufactured prior drawing improvement, the occurrence rate of this kind of event is (b)(4).Since the new design has been introduced, only one case has happened with an occurrence rate of (b)(4).Both these occurrence rates are overestimated because they do not consider the reuse of the instrument.Most of these issues happened after several uses of the resection jigs or of the pins.None of these cases caused serious consequences on the patient or prolonged surgical time.Limacorporate will continue to monitor the market to promptly detect the possible recurrence of this issue and assess the effectiveness of the corrective action performed.
 
Event Description
Intra-operative issue occurred on (b)(6) 2019.During anatomic smr surgery, the pin got stuck into the smr anatomic resction jig code 9013.50.304 lot #15ab119.Cutting jig from another tray was available and used.Surgery was prolonged of less than 10 minutes.Event happened in (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMR - ANATOMIC RESECTION JIG
Type of Device
SMR - ANATOMIC RESECTION JIG (KWT, HSD)
Manufacturer (Section D)
LIMACORPORATE SPA
via nazionale 52
villanova di san daniele, udine 33038
IT  33038
MDR Report Key9217396
MDR Text Key219377236
Report Number3008021110-2019-00123
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
PMA/PMN Number
K100858
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number9013.50.304
Device Lot Number15AB119
Date Manufacturer Received10/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-