The initial reporter stated they received questionable results for two patient samples tested for ise assays on the cobas 8000 ise module.Of the two samples, one had discrepant results for ise indirect na, ci for gen.2.No incorrect results were reported outside of the laboratory.The sample was repeated on a different analyzer and the repeat results were believed to be correct.The sample initially resulted with an na value of 124 mmol/l, which repeated as 144 mmol/l.The sample initially resulted with a cl value of 112 mmol/l, which repeated as 97 mmol/l.The serial number of the cobas 8000 ise module was (b)(4).
|