(b)(4).The pro240 device, lot number 94599, was returned for evaluation and visually and functionally tested.The complaint was confirmed, it was found that the deployment cable had been welded during manufacturing, which prevented the clip from deploying.
|
It was reported that on (b)(6) 2019, a patient underwent an off-pump video-assisted thoracoscopic surgery for left atrial appendage management procedure.The surgeon placed the pro240 on the laa and when the orange tab was pulled, the tab got separated from the wires.Surgeon used a new pro240 that was successfully deployed on the laa.Procedure was delayed for 10 minutes.There was no adverse consequence to the patient.
|