PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problems Display or Visual Feedback Problem (1184); Pacing Problem (1439)
Patient Problems Bradycardia (1751); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/09/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer displayed a certificate renewal request while a patient was being paced with the analyzer.The patient experienced a heart rate of 30 beats per minute (bpm).The user was unable to update the security certificate at the time because no internet was available in the operating room.A different model of programmer was used to pace the patient.The programmer remains in use.No further patient complications were noted.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Upon secondary review the reported event of unable to connect to the programmer analyzer did not cause or contribute to death or serious injury.No patient complications have been reported as a result of this event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient experienced a heart rate of 30 beats per minute (bpm).An attempt was made to connect to the mobile programmer analyzer in order to provide pacing however the programmer displayed a security certificate renewal request.The user was unable to update the security certificate at the time as internet was not available in the operating room.A different model of programmer was used to pace the patient.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK SMARTSYNC BASE
Type of Device
ANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9234661
MDR Text Key163712510
Report Number3004593495-2019-01119
Device Sequence Number1
Product Code DTC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24970A
Device Catalogue Number24970A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/16/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-