Device evaluated by mfr: the wolverine cb mr, ous 10 mm x 2.50 mm device was returned for analysis.A visual examination identified that the balloon folds were relaxed.A microscopic examination identified a balloon longitudinal tear starting approximately 1mm from the distal end of the proximal markerband extending for a length of approximately 1 mm distally.This type of damage most likely occurred due to the interaction of the device (the blade) and the guidezilla guide catheter extension.The markerbands, blades and tip section of the device were visually and microscopically examined, and no issues were noted with the markerbands, blades or tip of the device that may have potentially contributed to the complaint incident.All blades were present and fully bonded to the balloon material.A visual and tactile examination found a shaft polymer extrusion kink at approximately 100 mm distal of the mid-shaft/hypotube bond.This type of damage is consistent with excessive force that could have been applied on the delivery system.A visual and tactile examination found no issues with the hypotube shaft of the device that may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
|