• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VERSAPOINT ELECTRODE UNKNOWN PRODUCT; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VERSAPOINT ELECTRODE UNKNOWN PRODUCT; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Catalog Number VERSAUNK
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fever (1858); Inflammation (1932); Pain (1994); Injury (2348); Not Applicable (3189)
Event Type  Injury  
Manufacturer Narrative
(b)(4).This report is related to a journal article; therefore, no product will be returned for analysis and the batch history records cannot be reviewed as the lot number has not been provided.Attempts are being made to obtain the additional information.To date no response has been provided.If further details are received at the later date a supplemental medwatch will be sent.Was the case discussed in this article previously reported to ethicon? if yes, please provide a complaint reference number.Does the surgeon believe that ethicon product (versapoint) involved caused and/or contributed to the post-operative complications described in the article? does the surgeon believe there was any deficiency with the ethicon product (versapoint) involved? citation: acta obstetricia et gynecologica scandinavica 93 (2014) 218¿220 / doi: 10.1111/aogs.12274.(b)(4).
 
Event Description
Title : thermal injury to the sigmoideum following hysteroscopic hydrothermal ablation the purpose of this case report was to discuss thermal injury to the sigmoideum following hysteroscopic hydrothermal ablation.This was regarding a 39-year-old woman with a surgical history including laparoscopic enucleation of an endometrioma underwent hydrothermal endometrial ablation because of menorrhagia and severe dysmenorrhea.Hysteroscopy was performed exposing a small polyp with benign appearance, size 5 mm, in the posterior wall of the uterus.The polyp was removed with versapoint (ethicon).Hydrothermal ablation was started under hysteroscopy guidance and the procedure was carried out without any technical difficulties.After surgery, the woman suffered from an expected degree of abdominal pain, which was sufficiently treated with peroral (oral) analgesia.In the evening, the day after surgery, she suddenly experienced increasing abdominal pain and diarrhea.During the night she developed fever and in the morning she was immediately admitted to the hospital.She suffered from high fever, diarrhea and general abdominal pain but no peritonitis.The laboratory blood test showed normal hemoglobin concentration and very high level of inflammatory markers.The woman was immediately started on intravenous antibiotics.The woman¿s general condition deteriorated resulting in a decision for laparotomy.The intraoperative findings included intra-abdominal fluid, a pathological pale area on the sigmoideum, 1 cm in diameter, suggestive of heat injury, but no perforation.On the posterior wall of the uterus there was an equivalent pale area but no perforation.Drainage was left intra-abdominally.The woman recovered slowly and the drainage was removed 2 days postoperatively.Possible explanation for the induced injury to the serosa of the sigmoideum is the removal of the small polyp before the hydrothermal ablation.The removal most probably caused a weakening and thinning of the uterus wall, facilitating heat transmission through the myometrium.The temperature will drop while going through the tissue, which is why damage occurred only to the serosa of the sigmoideum.
 
Manufacturer Narrative
Additional information was requested and following obtained: questions: was the case discussed in this article previously reported to ethicon? if yes, please provide a complaint reference number.Does the surgeon believe that ethicon product (versapoint) involved caused and/or contributed to the post-operative complications described in the article? does the surgeon believe there was any deficiency with the ethicon product (versapoint) involved? answers: versapoint is a reliable instrument to remove polyps with.There was nothing wrong with the specific article used in the case described in the report.However, the possible weakening of the uterine wall after removing a polyp with versapoint might increase the risk of heat transmission through the uterine wall or perforation of hot fluid if the endometrial ablation via circulating hot fluid is done in the same process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPOINT ELECTRODE UNKNOWN PRODUCT
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key9239892
MDR Text Key167436368
Report Number2210968-2019-89056
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
PMA/PMN Number
K040302
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup
Report Date 09/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberVERSAUNK
Was Device Available for Evaluation? No
Date Manufacturer Received11/07/2019
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age39 YR
-
-