The clinical investigation is still on-going.A follow-up report will be submitted once additional information is obtained and the clinical investigation is complete.(b)(4).Pneumothorax is the most common side effect associated with the zephyr valve treatment (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).Targeted lobar deflation likely causes inflation of the ipsilateral lobe, which can result in a tear of the already compromised parenchymal tissue of the emphysematous ipsilateral lobe, resulting in a pneumothorax (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).In the liberate study (ide clinical study used to support pma p180002's approval), 26.6% of the zephyr valve subjects experienced a pneumothorax in the treatment period (less than or equal to 45 days).These were managed using standard of care procedures as per previously published guidelines (valipour, arschang, et al.Respiration 87.6 (2014): 513-521).In 17.4% of the events, the pneumothorax resolved without any additional intervention with subjects under careful observation.In over half the events (56.5%), the pneumothorax was managed with a chest-tube only.An additional 13% of the events were managed with a chest-tube and the temporary removal of one or more valves, while another 13% of the events were managed with a chest-tube and removal of all the implanted valves.Upon successful resolution of the pneumothorax, removed valves can be replaced.Patients that experienced a pneumothorax experienced clinical benefits of the zephyr valve treatment that were similar to the benefits experienced by patients who did not have a pneumothorax.The zephyr ebv system ifu and pulmonx training program both specifically reference pneumothorax as a known side effect of this procedure and the published guidelines (valipour, arschang, et al."expert statement: pneumothorax associated with endoscopic valve therapy for emphysema-potential mechanisms, treatment algorithm, and case examples." respiration 87.6 (2014): 513-521).The reported event aligns with the experience observed in the liberate clinical study and is an expected side effect to the zephyr valve treatment.
|
On (b)(6) 2019 the patient had a bronchoscopic lung volume reduction procedure and zephyr valves were placed.The patient had negative cultures at the time of the bronchoscopy.The patient developed a pneumothorax about 6 hours after the procedure, and a pigtail chest tube was placed.The pneumothorax resolved within 72 hours.The patient developed an abscess 72 hours after the valve procedure on (b)(6) 2019.On (b)(6) 2019, one or two valves were removed to drain and address the abscess.
|
Multiple attempts have been made to collect additional information about this event.Unfortunately, no additional information has been provided from the site after several months of communication attempts.If information is received at a later date, follow-up reports will be submitted.
|