Model Number MC2-4570S
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problems Failure of Implant (1924); Low Oxygen Saturation (2477); Blood Loss (2597)
Event Date 10/02/2019
Event Type  malfunction  
Manufacturer Narrative
No product has been returned for evaluation as it remains in-situ.Radiographs provided confirmed the alleged event.No root cause can be confirmed at this time.
 
Event Description
Information was received that during fluoroscopy it was observed that the end cap was disengaged from the rod.As per reporter, a revision procedure is planned.
 
Event Description
Received information stating the revision procedure was extended due to surgical complications.Patient was taken to the intensive care unit (icu) instead of general ward.
 
Manufacturer Narrative
Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Event Description
Additional information provided: d4, h6, h7 h9, h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
MDR Report Key9246070
MDR Text Key167994760
Report Number3006179046-2019-00168
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00812258022600
UDI-Public812258022600
Combination Product (y/n)N
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMC2-4570S
Device Catalogue NumberPA0684-007
Device Lot Number8061809
Was the Report Sent to FDA? No
Date Manufacturer Received11/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1898-2020
Patient Sequence Number1
Treatment
STRYKER XIA SCREW SYSTEM.
Patient Outcome(s) Hospitalization; Other;
Patient Age7 YR
-
-