Problem/evaluation code 213: the device was returned to w.L.Gore & associates for investigation.The following observations were made: the endoprosthesis, delivery catheter, part of the deployment line, and an introducer sheath were returned.The introducer sheath was not evaluated as it is not a gore product.The deployment line appeared to be broken.There was approximately 1.7cm still attached to the endoprosthesis with two single fibers coming from the end measuring approximately 0.5cm each.Approximately 0.7cm of the distal shaft, upon which the endoprosthesis is mounted, was exposed at the transition.Approximately 1.5cm of the endoprosthesis was partially expanded towards the tip end of the device.The remainder of the endoprosthesis was constrained by the inner braided constraining line.The endoprosthesis appeared to be damaged and contained outwardly flared struts and tape delamination throughout the expanded section.Deployment was able to be continued with traction on the deployment line at the endoprosthesis.The device history file was evaluated.No anomalies or non-routine maintenance were identified that could contribute to deployment line and constraint sleeve damage during assembly.Therefore, based on the evaluation performed, no manufacturing cause could be identified.
|