• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEWDEAL SAS MECHE NON CANULEE DIA 1,9; HALLUFIX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEWDEAL SAS MECHE NON CANULEE DIA 1,9; HALLUFIX Back to Search Results
Catalog Number 119618ND
Device Problem Dull, Blunt (2407)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/13/2019
Event Type  malfunction  
Manufacturer Narrative
To date, the device involved in the reported incident has not been received for evaluation.An investigation has been initiated based on the reported information.
 
Event Description
It was reported that the hallufix drill was not enough sharp.Product was in contact with the patient; however, no patient injury was reported and the event did not lead to surgical delay.1 of 2 reports: other mfg report number: 9615741-2019-00018.
 
Manufacturer Narrative
Additional information: h2, h3, h11 corrected information: b5, d10, g7, h6, h10.Sample was received for evaluation.Manufacturing records were reviewed and found no anomalies.Failure analysis: not confirmed, product is within specifications after inspection.Root cause: as no anomaly was found during documentary review and failure of the returned part, the issue cannot be duplicated.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa (b)(4) have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(4) office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MECHE NON CANULEE DIA 1,9
Type of Device
HALLUFIX
Manufacturer (Section D)
NEWDEAL SAS
97 allee alexandre borodine
97 allee alexandre borodine
saint priest 69800
FR  69800
MDR Report Key9261987
MDR Text Key166539036
Report Number9615741-2019-00019
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number119618ND
Device Lot NumberEDF7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2019
Date Manufacturer Received12/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-