• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA YORK, PA INC. MLX 300W XENON LIGHTSOURCE; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA YORK, PA INC. MLX 300W XENON LIGHTSOURCE; N/A Back to Search Results
Catalog Number 00MLX
Device Problems Melted (1385); Arcing (2583); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A customer reported that one month ago (unspecified date), the 00mlx mlx 300w xenon lightsource, the internal lamp module does not turn on when powered up.The ac adapter sends out an arc of electricity when connected to power and has caused some burning and melting of the plastic in the receptacle.A male patient was prepped for surgery however, there was no patient contact, injury or surgical delay reported.Request for additional information has been sent.
 
Manufacturer Narrative
The mlx light source was returned for evaluation.There was no external visible, physical damage observed.There was a basic lamp failure/malfunction.The reported complaint was confirmed from the evaluation.The underlying root cause for the reported event is undetermined: no manufacturing, workmanship, or material deficiency has been identified.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa pr 229048 and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(6) , director of regulatory programs, office of product evaluation and quality and (b)(6) , assistant director, mdr team, office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MLX 300W XENON LIGHTSOURCE
Type of Device
N/A
Manufacturer (Section D)
INTEGRA YORK, PA INC.
589 davies drive
589 davies drive
york PA 17402
MDR Report Key9267918
MDR Text Key166398095
Report Number2523190-2019-00131
Device Sequence Number1
Product Code FCW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00MLX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2019
Date Manufacturer Received12/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-