WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990739
Device Problems Device Displays Incorrect Message (2591); Failure to Fire (2610)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2019
Event Type  malfunction  
Manufacturer Narrative
The device history records (dhr) for the device were reviewed.The associated device was released based on acceptance criteria.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device was out of production prior to the september 24, 2014 udi regulation date.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported the laser stopped firing at 5% complete and displayed an energy our of range system message.Tech support was contacted and the system was restarted and performed a gas exchange.An energy check was done but a sensor misaligned message displayed.The procedure was cancelled.
 
Manufacturer Narrative
During on site visit, the fse (filed service engineer) found out that no output is produced from the laserhead after fresh fill, checked if e1 is blocked, and installed new laserhead.Performed system verification.Replaced part return has been requested but not provided.The root cause cannot be identified conclusively without sample.The most possible root cause is a defect of the laser head.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key9272274
MDR Text Key190823158
Report Number3003288808-2019-01083
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age33 YR
-
-