During the returned device evaluation, a distal bond peel was observed, but remained intact to the device.Recurrence of this malfunction could result in an embolism, myocardial ischemia or infarction, or prolonged procedure.Patient information regarding relevant tests/laboratory data, or medical history is unknown.This information was not available from the facility.The angiosculpt device was returned for evaluation.Visual inspection found a distal bond peel with five leaflets lifted, but remained intact to the device.In addition, the intermediate shaft was flattened and stretched, which measured approximately 14 cm over the specification.No defect was noted on the hypotube.Based on the lab analysis, it is likely excessive exertion of force was applied by the user, resulting in the damage observed on the device.Per the ifu, retained device component is listed as a possible adverse effect of the procedure.
|
During use in patient, the angiosculpt device was defective.The hypotube was stretched out and became floppy.The procedure was completed with a new device and no patient injury reported.During the returned device evaluation on 10/10/2019, a distal bond peel was observed.
|