The device history record for the affected lot was reviewed.No abnormalities that could have contributed to this event were found and the product was released according to manufacturer's acceptance criteria.The manufacturer performs a 100% final inspection for this product.The received information indicates that the device was reused.This is a off-label use and is considered as user error.The manufactured product is provided sterile and is intended for single use.The instrument is a sterile single use device and shall not be reused.Potential risk from reuse or reprocessing include: reduced cutting or grasping performance, and foreign particle introduction into the eye.The packaging indicates also that this device should not been reused.Reuse of a non reusable device is a use error.The dfu was not followed.(b)(4).
|