SMITH & NEPHEW, INC. HIP IMPLANT; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL Back to Search Results
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Injury (2348)
Event Date 06/01/2017
Event Type  Injury  
Event Description
It was reported that the patient presented a right hip dislocation.The event was treated via closed reduction.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that no relevant clinical information has been provided to perform a thorough medical investigation.Should any additional medical information be provided this compliant will be reassessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.Some potential probable causes of this event could include improper surgical technique or device selection.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP IMPLANT
Type of Device
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9279849
MDR Text Key165085121
Report Number1020279-2019-03876
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study,use
Type of Report Initial,Followup
Report Date 02/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received02/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-