MAUDE Adverse Event Report: ZIMMER GMBH METASUL LDH, HEAD, 46, CODE L, TAPER 18/20; N/A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH METASUL LDH, HEAD, 46, CODE L, TAPER 18/20; N/A Back to Search Results
Model Number N/A
Device Problem Material Erosion (1214)
Patient Problems Unspecified Infection (1930); Osteolysis (2377); Reaction (2414)
Event Date 10/14/2019
Event Type  Injury  
Manufacturer Narrative
Medical products and therapy date detail of product: item number 0100214052, item name metasul durom, component for acetabulum, uncemented, 52¸ 46, code l lot # 2267929.Item number 0100185146, item name metasul ldh, head adapter, m, 0, taper 12/14- 18/20, lot # 2291567.Item number 290039080, item name cls spotorno, stem, 135, uncemented, 8.0, taper 12/14, lot # 2278300.The manufacturer did not receive devices or x-rays for review.Where lot numbers were received for the devices, the device history records were reviewed and found to be conforming.Based on an extensive investigation of events reported from several user facilities outside the usa, zimmer identified that the most probable cause for the outcome observed was a loose or unstable cup that resulted from use of surgical techniques not consistent with the manufacturer's recommendations.As a corrective action, a retraining program for users outside the usa was initiated in november 2009 and reported to the national competent authorities as required.The durom cup reported in this case is not marketed in the usa.A similar cup, compatible with the metasul ldh femoral head, is cleared in the usa and a corrective action for this product was reported to the fda in july 2008 as notification z-2415/2426-2008.No further investigation required as this issue is known and addressed in cp00000620 (error pattern: potential early revision of the acetabular component due to loosening, implant migration or unresolved pain, higher ion release).At least one of these error patterns is observed in this event.Should any additional information that changes the assessment become available to us, since this case is related to the issues for which zimmer implemented a corrective action there will be no further investigation.Therefore, zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
 
Event Description
It was reported that patient underwent revision surgery due to periprosthetic infection, osteolysis and elevated metal ions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METASUL LDH, HEAD, 46, CODE L, TAPER 18/20
Type of Device
N/A
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key9283672
MDR Text Key165262855
Report Number0009613350-2019-00656
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/15/2010
Device Model NumberN/A
Device Catalogue Number01.00181.460
Device Lot Number2276764
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/28/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age47 YR
-
-