Model Number 3854
Device Problem Break (1069)
Patient Problem No Code Available (3191)
Event Date 10/03/2019
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: over 18 years old.
 
Event Description
It was reported that the catheter shaft break occurred.The target lesion was located in the moderately tortuous and severely calcified mid left anterior descending artery.A 10mmx3.0mm wolverine coronary cutting balloon was selected for use.During procedure, the balloon was advanced into the lesion.However, the balloon catheter shaft broke into half before inflation.The proximal part of the balloon catheter was then removed and a 2.5mmx16mm emerge balloon was used.Subsequently, the emerge balloon was inflated and trapped the remaining wolverine shaft inside the guide catheter and facilitated the removal of all the device component from the patient's body.The procedure was completed with a new guide catheter and a non-compliant balloon.No complications reported and patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9284051
MDR Text Key168072266
Report Number2134265-2019-13272
Device Sequence Number1
Product Code NWX
UDI-Device Identifier08714729888260
UDI-Public08714729888260
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2019
Device Model Number3854
Device Catalogue Number3854
Device Lot Number0021388752
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-