Catalog Number UNK FEMORAL STEM TRIAL
Device Problems Incorrect Measurement (1383); Appropriate Term/Code Not Available (3191)
Patient Problem Bone Fracture(s) (1870)
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The trial necks can be impossible to seat at times due to medial bone when the broach is below the level of the neck osteotomy.There is an option to remove this bone with the reamer, but that then commits the surgeon to that length of neck cut which when using a collared implant may be inappropriate.Again i am having made locally some neck trials that have the medial portion of the trial removed.This then allows the neck trial to be used even when the broach is well below the neck osteotomy.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK FEMORAL STEM TRIAL
Type of Device
UNK FEMORAL STEM TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9286175
MDR Text Key166982223
Report Number1818910-2019-114489
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK FEMORAL STEM TRIAL
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received01/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-