MAUDE Adverse Event Report: BIOMERIEUX, SA VITEK® MS; VITEK MS INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMERIEUX, SA VITEK® MS; VITEK MS INSTRUMENT Back to Search Results
Model Number 410895
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
A customer from (b)(6) notified biomérieux of a misidentification of cutibacterium acnes while using the vitek® ms instrument (ref.410895, serial number (b)(4)).The customer had both an aerobic and anaerobic blood culture bottle of the same patient sample.Both blood culture bottles were tested on the vitek® ms instrument after sub-culturing on agar plates.The vitek ms instrument obtained an identification of cutibacterium acnes for the aerobic culture and vibrio fluvialis or vibrio fluvialis/ e.Faecalis for the anaerobic culture.Microscopic examination showed an identification of cutibacterium acnes for both cultures.Due to the discrepant result, the customer repeated the testing on the anaerobic culture on the same day.Again an identification result of vibrio fluvialis was obtained.On (b)(6) 2019, both blood culture isolates were tested again, and the vitek® ms instrument gave identification results of cutibacterium acnes for both bottles.Based on microscopic examination and the vitek® ms result, the identification of cutibacterium acnes was reported, and patient treatment was not affected.There was no adverse event related to the patient's state of health as a result of this misidentification.A biomérieux internal investigation will be initiated.
 
Manufacturer Narrative
This report was initially submitted when a customer from germany notified biomérieux of a misidentification of cutibacterium acnes (propionibacterium acnes) as vibrio fluvialis or vibrio fluvialis/ e.Faecalis using the vitek® ms instrument (ref.410895, serial number (b)(6).An internal biomérieux investigation was performed analyzing the data provided by the customer.Conclusion on the fine tuning of the vitek ms instrument: the fine tuning done before the identification issue was non-optimal.A new fine tuning was done on (b)(6) 2019.The analysis of the calibrator mzml files indicated that a fine tuning was needed during the tests performed on (b)(6) 2019.Conclusion on spot preparation quality: the customer's spot preparation quality was non-optimal.The calibrator and sample "all peaks" values are heterogeneous.This could be explained by a non-optimal spot preparation of the aample and calibrator strains (culture, spot, different operator.).Conclusion on the identification: with the information provided, the expected identification is cutibacterium acnes (= propionibacterium acnes).The identification could be confirmed with the reference method (sequencing).For information, the misidentification was obtained from the spectra having the fewer number of peaks.Moreover, the analysis of the sample mzml files indicates that the quality of the sample preparation was non-optimal.It could explain the misidentification obtained.Suspected causes retained: non-optimal fine tuning.Lack of fine tuning monitoring.Non-optimal spot preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITEK® MS
Type of Device
VITEK MS INSTRUMENT
Manufacturer (Section D)
BIOMERIEUX, SA
3 route de port michaud
la balme, 38390
FR  38390
MDR Report Key9292783
MDR Text Key220259881
Report Number9615754-2019-00117
Device Sequence Number1
Product Code PEX
UDI-Device Identifier03573026359119
UDI-Public03573026359119
Combination Product (y/n)N
PMA/PMN Number
DEN130013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number410895
Was Device Available for Evaluation? No
Date Manufacturer Received01/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-